Your Cart (0)
IPAQ R330 DIN RAIL Mounted Universal Programmable 2-wire 4 to 20mA Temperature Transmitter
Overview
IPAQ C330 is a modern temperature transmitter developed to meet the highest requirements on flexibility, accuracy and reliability. A universal transmitter compatible with RTD, thermocouples, voltage and potentiometer sensors to offer high flexibility and reduce the number of installed product variants. IPAQ C330 is available with ATEX, IECEx and cFMus certifications, making it a safe choice for a wide range of applications. The IPAQ C330 is scalable over the entire range of 7 types of RTDs and 10 Thermocouple Types; as well as accepting Millivolt and Resistance inputs.Features
- High accuracy and long term stability
- Universal Input - Accepts RTD, T/C, mV and Ω
- 50-point Customized Linearization and Callendar-Van Dusen/li>
- Sensor error and system (sensor/transmitter) error correction for highest total accuracy
- Low temperature drift
- Configuration via USB or NFC without external power
- Rugged design tested for 10 g vibrations
- High security - Password protection and date of changes logged
Price Starting At:
$208.00
$208.00
Shipping Estimated 1 week.
Click Add to Cart to Select Temperature Sensor, Scale and Range
Please add 1 Week to Estimated Shipping for Configuration
Click Add to Cart to Select Temperature Sensor, Scale and Range
Please add 1 Week to Estimated Shipping for Configuration
IPAQ R330 Programmable Transmitter Specifications
Input Types | Thermocouple - J, K, T, E, N, R, S, B, C, D RTD - Pt100 (IEC 60751, α = 0.00385) AND (JIS C1604-1981, α = 0.003916), Ni100, Ni120, Ni1000, Cu10, resistance / potentiometer |
Sensor Range, Accuracy, Minimum Span | See Range and Accuracy Table Below |
Output | 4-20 mA / 20-4 mA, temperature linear |
Operating temperature | -40 to +185 °F (-40 to +85°C) |
Galvanic isolation | 1500 VAC, 1 min |
Power Supply | IPAQ C330 8.0 to 36.0 VDC IPAQ C330X 8.0 to 30.0 VDC |
Measuring Accuracy | Accuracy & stability - Basic accuracy is max. of ±0.08°C or ±0.08% of span |
Ambient Temperature Influence | Thermocouple and RTD: See Table Below Resistance - ±0.01% < 4000 Ω (2000 Ω at 2-wire) < ±0.02% of span per °C Voltage: ±0.01% of span per °C |
Supply Voltage Influence | <±0.005% of span per V | Long Term Drift | Max. of ±0.02°C or ±0.02% of span per year | Operating and Storage Temperature | Standard version: -40 to +185°F (-40 to +85°C) Intrinsically safe transmitters Temperature Class T6 - Ambient temperature Ta -40°C ≤ Ta ≤ +60°C / -40°F ≤ Ta ≤ +140°F Temperature Class T5 - Ambient temperature Ta -40°C ≤ Ta ≤ +75°C / -40°F ≤ Ta ≤ +167°F Temperature Class T4 - Ambient temperature Ta -40°C ≤ Ta ≤ +85°C / -40°F ≤ Ta ≤ +185°F |
Humidity | 0 to 98% RH (non-condensing) | Protection Category | Housing: IP65 Terminals: IP00 |
Housing Materials | Jouding PC/ABS + PA Flammability according to UL VO |
Thermocouple Range, Accuracy and Minimum Span Table
Sensor Type | Temperature Range °F/°C | Minimum Span °F/°C | Accuracy (maximum of) °F/°C | Temp. influence (Dev. from ref. temp. 20°C) °F/°C |
---|---|---|---|---|
Type K (NiCr-NiAl) | -454 to +2372°F -270.to.+1300°C |
90°F 50°C |
±0.9°F or ±0.1% of span (2)*
±0.5°C or ±0.1% of span (2)* |
±0.006% of span per °F
±0.01% of span per °C |
Type J (Fe-CuNi) | -346 to +2192°F -210.to.+1200°C |
90°F 50°C |
±0.9°F or ±0.1% of span (2)* ±0.5°C or ±0.1% of span (2)* |
±0.006% of span per °F ±0.01% of span per °C |
Type T (Cu-CuNi) | -454 to +752°F
-270 to +400°C |
90°F 50°C |
±0.9°F or ±0.1% of span (2)* ±0.25°C or ±0.2% of span (2)* |
±0.006% of span per °F ±0.01% of span per °C |
Type E (NiCr-CuNi) | -454 to +1832 °F -270 to +1000 °C |
90°F 50°C |
±0.9°F or ±0.1% of span 2 °F (2)* ±0.5°C or ±0.1% of span °C (2)* |
±0.01% of span per °C |
Type N (NiCrSi-NiSiMg) | 148 to +2372°F -100 to +1300°C |
<180°F 100°C |
±0.9°F or ±0.1% of span (2)* ±0.5°C or ±0.1% of span (2)* |
±0.006% of span per F° ±0.01% of span per °C |
Type N (NiCrSi-NiSiMg) | -454 to -148 F° -270 to -100°C |
180°F 100°C |
±1.8°F (2)* ±1.0 °C(2)* | ±0.18% of span per °F ±1.0°C or ±0.1% of span (2)* |
S (Pt10%Rh-Pt) | -58 to +3182F° -50 to +1750°C |
540°F 300°C |
±1.8°F or ±0.1% of span 2 ±1.0°C or ±0.1% of span (2)* |
±0.006% of span per °F ±0.1% of span per °C |
R (Pt13%Rh-Pt) | -58 to +3182F° -50 to+1750°C |
540°F 300°C |
±1.8°F or ±0.1% of span 2 ±1.0°C or ±0.1% of span (2)* |
±0.006% of span per °F ±0.1% of span per °C |
B (Pt30%Rh-Pt6%Rh) | +752 to +3272F° +400 to +1800C° |
1260F° 700°C |
±1.8°F or ±0.1% of span (2)* ±1.0°C or ±0.1% of span (2)* |
±0.006% of span per °F ±0.1% of span per °C |
C (W5%Re-W26%Re) | +32 to +4199°F (2)* 0 to +2315°C |
360F° 200°C |
±1.8°F or ±0.1% of span (2)* ±1.0°C or ±0.2% of span (2)* |
±0.006% of span per °F ±0.1% of span per °C |
D (W3%Re-W25%Re) | +32 to +4199°F (2)* 0 to +2315°C |
360F° 200°C |
±1.8°F or ±0.1% of span (2)* ±1.0°C or ±0.2% of span (2)* |
±0.006% of span per °F ±0.1% of span per °C |
*Note 2 CJC error not included. = 0.5°C within ambient temperature range
RTD Range, Accuracy and Minimum Span Table
Sensor Type | Temperature Range °F/°C | Minimum Span °F/°C | Accuracy (maximum of) °F/°C | Temp. influence (Dev. from ref. temp. 68°F/20°C) °F/°C |
---|---|---|---|---|
RTD Pt100 (a =0.00385) (1)* | -328 to +1562°F -200 to +850°C |
18°F 10°C |
±0.18°F or ±0.1% of span ±0.08°C or ±0.08% of span |
±0.006% of span per °F ±0.1% of span per °C |
RTD Pt100 (a=0.003916) (2)* | -328 to +1562°F -200 to +850°C |
18°F 10°C |
±0.18°F or ±0.1% of span ±0.08°C or ±0.08% of span |
±0.006% of span per °F ±0.1% of span per °C |
RTD PtX (1)* | Corresp. to max.4 k | 18°F 10°C |
±0.18°F or ±0.1% of span ±.1°C or ±0.1% of span |
±0.006% of span per°F ±0.01% of span per °C |
RTD Ni100 (3)* | -76 to +482°F -60.to.+250°C |
18°F 10°C |
±0.18°F or ±0.1% of span ±0.1°C or ±0.1% of span |
±0.006% of span per °F ±0.01% of span per °C |
RTD Ni120 (4)* | -76 to +482°F -60.to.+250°C |
18°F 10°C |
±0.18°F or ±0.1% of span ±0.1°C or ±0.1% of span |
±0.006% of span per °F ±0.01% of span per °C |
RTD Ni1000 (3)* | -58 to +356°F -50.to.+180 |
18°F 10°C |
±0.18°F or ±0.1% of span ±0.1°C or ±0.1% of span |
±0.006% of span per °F ±0.01% of span per °C |
RTD Cu10 (Edison Copper Windings No. 15) | -58 to +392°F -50.to.+250°C |
149°F 83°C |
±2.7°F or ±0.2% of span ±1.5°C or ±0.2% of span |
±0.006% of span per °F ±0.02% of span per °C |
Sensor Current | Sensor current ≤300 μA |
Maximum sensor wire resistance | 3-wire and 4-wire connection 50 Ω/wire 2-wire connection Compensation for 0 to 100 Ω loop resistance |
Adjustment | Minimum span 10°C / 18°F Sensor error compensation ±10% of span for span <50°C / 90°F, otherwise ±5°C/±9°F |
Intrinsic Safety Approvals
PART NUMBER | DESCRIPTION | IS RATINGS |
---|---|---|
70R330X012 | IPAQ R330X (ATEX, IECEx) | II 1 G Ex ia IIC T6…T4 Ga |
70R330F012 | IPAQ R330X (cFMus, ATEX, IECEx) | Ex ia IIC T6…T4 Ga |
70R330F012 | IPAQ R330X (cFMus, ATEX, IECEx) IPAQ R330X FM CA |
IS CL I Div 1 GP A-D, T6…T46) Cl I Zn 0 AEx/Ex ia IIC T6…T4 Ga6) |