Your Cart (0)
AMI 310 Standard Multifunction Meter Kit
Overview
AMI 310 PRO Multifunction Meter Kit Measures Pressure, Air Velocity, Air Flow, Temperature, Humidity, Dew Point. Each AMI 310 PRO Kit includes, color backlit graphic display, with 2 input channels for measuring probes and PT100 temperature. probes (-200 to +600°c), 2 thermocouple K, J, T, S input channels (-200 +1760°C), output for PC and printer.Sensing Probes and Modules Included
- Pressure Module Range ±500 pa and 2 to 28 m/s
- Telescopic Hotwire Probe, 1m length, with graduation, swiveling at 90° (0.15 to 30 m/s)
- Stainless Steel Hygrometry probe (3 to 98 %RH and -40 +180°C)
- 100 mm telescopic vane probe, length 1m, swiveling at ± 90° (0.3 to 35 m/s)
- Pitot Tube length 300 mm, 6 mm with integrated K Thermocouple probe
- 2 x 1 m of silicone tube (black and white)
- stainless steel tube ø 6mm length 100mm
- 2 cables with mini-din connector, for measuring probes
AMI 310 Standard Multifunctional Meter Specifications
Measuring Parameters | Air Velocity, Airflow, Differential Pressure, Humidity, Temperature |
Measuring Units (Dependent of Sensor Type) | °F, °C, % RH, Pa, rpm, m³/s, mph, kPa, ppm, g/m³, g/kg, m/s, bar, mbar, l/s, hPa, m³/h, tr/min, km/h, fpm, inWg, mmH2O, mmHg, daPa, cfm, psi |
Instrument Connections | Top: 2 secured mini-DIN connectors for SMART-Plus probes Bottom: 1 micro-USB port for charging and PC connection |
Connections Included Probes/Modules | MPR 10000 Pressure Module SHR 110 ABS hygrometry probe SSFC 300 Hotwire probe SH 100 4" (100 mm) diameter vane probe Interchangeable probes/modules have the SMART-2014 system and are automatically recognized once the connection is made on the instrument Optional Probes and Modules Sold Seperately |
Data Logging | Up to 1000 dataset of 20,000 points in the internal memory plus 4 GB micro-SD card (Included) |
Display | Graphic display 120 x 160 pixels Dim. 2.28" x 2.99" (58 x 76 mm), color Display of 6 measurements including 3 simultaneously |
Keypad | 3 Function Keys, Navigation Arrors, On/Off and ESC Key |
Conformity | 2014/30/EU ECM; 2014/35/EU Low tension; 2011/65/UE RoHS II; 2012/19/UE DEEE |
Power Supply | Rechargeable Lithium-Ion battery |
Battery Life | Approximately 32 hours with hygrometry probe |
Operating Environment | Neutral gas |
Operating Temperature | 32 to 122°F (0 to 50°C) |
Storage Temperature | 0 to 176°F (-20 to +80°C) |
Auto Shut-off | Adjustable from 15 to 120 min |
Weight (Meter Only) | 1.1 lb. (485g) |
Standard Languages Included | English, Spanish, Portuguese Optional Languages Please Call For Avaiabilty French, Dutch, German, Italian, Swedish, Norwegian, Finn, Danish |
AMI 310 STD Multi-function Meter Included Probes and Modules
MPR-10000 Pressure/Temperature ModuleMeasuring Parameter | Measuring Units | Measuring Range | Measuring Accuracy | Measuring Resolution | Tolerated Overpressure | Pressure | PPa, mmH2O, inWg, mbar, hPa, mmHg, daPa, kPa | 0 to +10,000 Pa or 0 to -10,000 (0 to ±40.19 inWg) | ±0.2% of reading ±10 Pa ±0.2% of reading ±0.04 inWga |
(1 Pa (0.004 inWg) |
1,200 mbar (481 inWg) |
---|
Measuring Parameter | Measuring Units | Thermocouple Type and Range | Measuring Accuracy | Measuring Resolution |
---|---|---|---|---|
Temperature | °C, °F | Type K: -328 to 2372°F (-200 to +1300°C) Type J: -148 to 1382°F (-100 to +750°C) Type T: -328 to 752°F (-200 to +400°C) Type N: -328 to 2372°F (-200 to 1300°C) Type S: 32 to 3200°F (0 to 1760°C) |
Types K, J, T, N: -328 to 32°F (-200 to 0°C): ±0.4°C ±0.3% of reading. 32 to 2372°F (0 to 1300°C)±0.8 °F (±0.4 °C) Type S: ±1.1 °F (±0.6 °C) |
|
Measuring Parameter | Measuring Units | Measuring Range | Measuring Accuracy | Measuring Resolution | Air Velocity | m/s, fpm, km/h, mph | -984 to 590.5 fpm (-5 to 3 m/s) 610 to 6,890 fpm (3.1 to 35 m/s) |
60 to 590 fpm: ±3% of reading±20 fpm (0.3 to 3 m/s: ±3% of reading ±0.1m/s) 591 to 6,890 fpm: ±1% of reading ±60 fpm (3.1 to 35 m/s: ±1% of reading ±0.3 m/s) |
0.1 fpm 0.01 m/s 1 fpm 0.1 m/s |
---|---|---|---|---|
Air Flow | m3/h, cfm, l/s, m3/s | 0 to 99,999 cfm (0 to 99,999 m3/h) |
±3% of reading ±0.00003*sheath surface in ft2 (±3% of reading or ±0.03*sheath surface (cm2) |
1 cfm 1 m3/h |
Temperature | °C or °F | -4 to 176°F (-20 to 80°C) | ±0.4% of reading ±0.3°C | 0.1°C |
Measuring Parameter | Measuring Units | Measuring Range | Measuring Accuracy | Measuring Resolution | Air Velocity | m/s, fpm, km/h, mph | 30 to 590 fpm (0.15 to 3 m/s) 591 to 5,900 fpm (3.1 to 30 m/s) |
±3% of reading ±6 fpm (±3% of reading ±0.03 m/s) ±3% of reading ±20 fpm (±3% of reading ±0.1 m/s) |
1 cfm (0.01 m/s) 0.1 m/s |
---|---|---|---|---|
Air Flow | m,3/h, cfm, l/s, m3/s | From 0 to 99,999 m3/h | ±3% of reading or ±0.03*sheath surface (cm2) | 1 m3/h |
Temperature | °C or °F | -4 to 176°F (-20 to 80°C) | ±0.3% of reading ±0.25°C | 0.1°C |
Measuring Parameter | Measuring Units | Measuring Range | Measuring Accuracy | Measuring Resolution | Relative Humidity | % RH | 0 to 100% RH | Accuracy (Repeatability, linearity, Hysteresis) ±1.5% RH (from 15°C to 25°C and from 3 to 98% RH) Factory calibration uncertainty: ±0.88% RH Temperature dependence: ±0.04 x (T-20) % RH (if T<15°C or T>25°C) |
0.1% RH |
---|---|---|---|---|
Absolute Humidity | g/m3 | From 0 to 600 g/m3 | 0.1 g/m3 | |
Dew Point | °Ctd, °Ftd | -58 to 212°Ftd (-50 to +100°Ctd) | ±0.6% of reading ±0.5°Ctd | 0.1 °C |
Wet Temperature | °C, °F | -58 to 212°F (-50 to +100°C) | 0.1 °C | |
Temperature | °C, °F | -4 to 176°F, (-40 to +180°C) | ±0.3% of reading ±0.25°C | 0.1 °C |
Each AMI 310 STD Standard Multi-function portable instrument is supplied with, with colour backlit graphic display, with 2 input channels for measuring probes and PT100 temperature probes (-200 to +600°C), 2 Thermocouple K, J, T, S input channels (-200 to + 1760°C), output for PC and printer. Supplied with MPR 10000 pressure module ±10000, SFC 300hot wire probe, SHR 110 ABSL hygrometry probe (3 to 98 % RH and -40 +180°C), SH 100 4" (100 mm) 100mm vane probe, 2 x 1 m of silicone tube (black and white), stainless steel tube 6mm length 100mm, 2cables with mini-DIN connector, for measuring probes, 2 lithium ion rechargeable batteries, with USB/mains adapter cable, Micro-SD board for data storage, calibration certificate and transport case.