Log In        Register

TC-100 Thermocouple, millivolt, mV, hand held Calibrator,Martel

TC100 Thermocouple/mV Hand Held Calibrator
    Price Starting At: $867.00

    Overview

    The TC-100 Thermocouple/mV Calibrator provide high accuracy and measurement of 10 common thermocouples, as well as mV. The TC-100's accuracy of ±0.3°C for Type J T/Cs includes all errors, at resolutions of 0.01°C or °F in measure mode, and 0.1°C or °F in source mode. Features including MIN/MAX recall in measure mode, three setpoints per thermocouple range, a large knob for decade control of the output in source mode, and the ability to accept bare Thermocouple wires in addition to mini-plug inputs, makes the TC100 an accurate, easy-to-use instrument for all your thermocouple calibration needs.
    Features
    • High accuracy ±0.3°C (Type J T/C all errors combined)
    • Ten (10) common T/C types plus mV
    • Accepts both T/C mini-plug and bare T/C wires
    • Simple decade control of output
    • Three (3) setpoints for each Thermocouple type
    • MIN/MAX recall in measure mode
    • Input protection to 240 VAC
      Shipping Please Call for availability
    914-673-0333

    TC-100 General Specifications
    Operating Temperature -10°C to +50°C (-14°F to 122°F)
    Storage Temperature -20°C to +70°C (-4°F to 158°F)
    Power Requirements 9 VDC. 9 V alkaline: NEDA1604 Optional NiCad Optional AC adapter/charger
    CJC Temperature Offset ±0.05°C per °C outside of 23 ±5°C
    Warm-up Time 1 minute to specification
    Dimensions 5.7" x 3.2" x 1.4" (H x W x D) 144.7 x 80 x 36.3 mm)
    Weight 12 ounces (340 grams)
    Other IP54
    Protected against misconnection to 250 VAC/VDC
    TC-100 Thermocouple Range and Accuracy Table
    Sensor Type Temperature Range Accuracy with CJC On
    Type J -210°C to -100°C (-364°F to -148°F)
    -100°C to +1200°C 9-148°F to +2192°F)
    ±0.5°C (±0.9°F)
    ±0.3°C (±0.5°F)
    Type K -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 1000.0°C (32 to 1832°F)
    1000.0 to 1372.0°C (1832 to 2498°F)
    ±0.6°C (±1.1°F)
    ±0.35°C (±0.63°F)
    ±0.5°C (±0.9°F)
    Type T -200°C to -150°C (-328°F to -302°F)
    -150°C to +400°C (-302°F to +752°F)
    ±0.7°C (±1.3°F)
    ±0.3°C (±0.5°F)
    Type E -200°C to -100°C; (-328°F to -148°F)
    -100°C to +1000°C (-148°F to +1832°F)
    ±0.5°C (±0.9°F)
    ±0.3°C (±0.6°F)
    Type R 0°C to 250°C (32°F to +482°F)
    250°C to +1767°C (+482°F to +3212°F)
    ±1.8°C (±3.2°F)
    ±1.0°C (±1.8°F)
    Type S 0°C to 250°C (32°F to +482°F)
    250°C to +1767°C (+482°F to +3212°F)
    ±1.8°C (±3.2°F)
    ±1.0°C (±1.8°F)
    Type B 600°C to 1000°C (+1112°F to +1832°F)
    1000°C to 1820°C (+1832°F to +3308°F)
    ±1.7°C (±3.1°F)
    ±1.2°C (±.2.2°F)
    Type C 0 to 1000°C (32 to 1832°F)
    1000 to 2318°C (1832 to 4204°F)
    0.8°C (1.4°F)
    2.5°C(4.5°F)
    Type L -200 to 0.0°C (-328 to 32°F)
    0.0 to 900.0°C (32 to 1652°F)
    0.5°C (0.9°F)
    0.4°C(0.7°F)
    Type U -200 to 0.0°C (-328 to 32°F)
    0.0 to 400.0°C (32 to 752°F)
    0.7°C (1.3°F)
    0.45°C (0.8°F)
    mV Read and Source -10.000mV - 75.000mV 0.02% ± 10µV
    Specifications (1 year at 23 °C ±5 °C; % of reading unless otherwise noted).
    Each instrument includes Protective Sleeve Case, Batteries, Instruction Manual, NIST Calibration Certificate.
    close