    The MC-1010 Multifunction Calibrator high accuracy with the New ClearBrite™ High Graphic Display. This single instrument will calibrate virtually any process device. Capable of reading and simulating Thermocouples, RTDs, Ohms, Voltage, Current and Frequency. The MC-1010 features a simple user interface, large display, and an RS232 interface.
    Features
    • Ten (10) T/C types and ten (10) RTD types, plus YSI 400 Thermistors
    • Read and source modes
    • ±0.4°C thermocouple/±0.3°C RTD accuracy
    • Full function 4 to 20mA read, source, simulate, and read/power modes with ±0.015% of reading accuracy
    • Source or read 0 to 20 Volts (±0.015% of reading accuracy)
    • Direct keyboard entry enables quick output changes
    • Store up to nine (9) setpoints for each output function
    • RTD function compatible with all pulsed transmitters
    • RS232 interface
    • New ClearBrite™ High Graphic Display
      Shipping Please Call for availability
    914-673-0333

    MC-1010 Multifunction Temperature and Process Calibrator Specifications
    ClearBriteTM Display High contrast 128 x 64 pixel addressable graphic LCD w/daylight backlight 2.4 x 1.8 in., 63 x 44 mm
    Serial Communications Yes, ASCII, RS-232, requires optional Martel 1919069 serial cable or 1919896 USB cable
    Operating Temperature -10°C to +50°C (-14°F to 122°F)
    Storage Temperature -20°C to +70°C (-4°F to 158°F)
    Power Requirements 6Vdc; Batteries 4 AA alkaline (supplied) or optional rechargeable
    CE - EMC EN50082-1 and EN55022: 1994 Class B
    Safety CSA C22.2 No. 1010-1: 1992
    Dimensions 8.5" H x 4.8" W x 2.1" D (22 x 12 x 5.3 cm)
    Weight 1.8 lbs. (0.82 kg)
    Other IP54
    Protected against misconnection to 250 VAC/VDC
    MC-1010 Thermocouple Range and Accuracy Tables
    Sensor Type Temperature Range Accuracy with CJC On
    Type J -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 800.0°C (32 to 1472°F)
    800.0 to 1200.0 (1472 to 2192°F)
    0.6°C (1.1°F)
    0.4°C (0.7°F)
    0.5°C (0.9°F)
    Type K -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 1000.0°C (32 to 1832°F)
    1000.0 to 1372.0°C (1832 to 2498°F)
    0.8°C (1.4°F)
    0.5°C (0.9°F)
    0.7°C (1.3°F)
    Type T -250.0 to 0.0°C (-418 to 32°F)
    0.0 to 400.0°C (32 to 752°F)
    0.8°C (1.4°F)
    0.5°C (0.9°F)
    Type E -250.0 to -100.0°C (-418 to -148°F)
    -100.0 to 1000.0°C (-148 to 1832°F)
    0.8°C (1.5°F)
    0.4°C (0.7°F)
    Type R 0 to 1767°C (32 to 3203°F) 1.4°C (2.5°F)
    Type S 0 to 1767°C (32 to 3203°F) 1.4°C (2.5°F)
    Type B 600 to 800°C (1112 to 1472°F)
    800 to 1000°C (1472 to 1832°F)
    1000 to 1800v (1832 to 3272°F)
    1°C (1.8°F)
    1.5°C (2.7°F)
    1.7°C (3°F)
    Type C 0 to 1000°C (32 to 1832°F)
    1000 to 2318°C (1832 to 4204°F)
    0.8°C (1.4°F)
    2.5°C(4.5°F)
    Type L -200 to 0.0°C (-328 to 32°F)
    0.0 to 900.0°C (32 to 1652°F)
    0.5°C (0.9°F)
    0.4°C(0.7°F)
    Type U -200 to 0.0°C (-328 to 32°F)
    0.0 to 400.0°C (32 to 752°F)
    0.7°C (1.3°F)
    0.45°C (0.8°F)
    Type N -200 to 0.0°C (-328 to 32°F)
    -0 to 600.0°C (32 to 1112°F)
    600.0 to 1300.0°C (1112 to 2372°F)
    1°C (1.8°F)
    0.5°C (.9°F)
    0.6°C (1.1°F)
    mV Read and Source -10.000mV - 75.000mV 0.02% ± 10μV
    Specifications (All specifications apply at 23 °C ± 5°C unless otherwise stated. Outside this range add ±0.005% of reading/°C. Accuracy is % of reading ± floor spec.)
    MC-1010 RTD Range and Accuracy Tables
    Sensor Type Temperature Range Accuracy
    Pt10 (385) -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 100.0°C (32 to 212°F)
    100.0 to 300.0°C (212 to 572°F)
    300.0 to 400.0°C (572 to 752°F)
    400.0 to 630.0°C (752 to 1166°F)
    630.0 to 800.0°C (1166 to 1472°F)
    1.3°C (2.3°F)
    1.3°C (2.3°F)
    1.4°C (2.5°F)
    1.5°C (2.7°F)
    1.8°C (2.9°F)
    1.9°C (3.4°F)
    Pt50 (385) -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 100.0°C (32 to 212°F)
    100.0 to 300.0°C (212 to 572°F)
    300.0 to 400.0°C (572 to 752°F)
    400.0 to 630.0°C (752 to 1166°F)
    630.0 to 800.0°C (1166 to 1472°F)
    0.4°C (0.72°F)
    0.4°C (0.72°F)
    0.4°C (0.72°F)
    0.5°C (0.9°F)
    0.5°C (0.9°F)
    0.6°C (1.8°F)
    Pt100 (385) -200.0 to -80.0°C (-328 to -112°F)
    -80.0 to 300.0°C (-112 to 572°F)
    300.0 to 630.0°C (572 to 1166°F)
    630.0 to 800.0°C (-328 to 32°F)
    0.1°C (0.18°F)
    0.2°C (0.36°F)
    0.3°C (0.54°F)
    >0.6°C (1.8°F)
    Pt200 (385) -200.0 400.0°C (-328 to 752°F)
    400.0 to 630.0°C (752 to 1166°F)
    0.7°C (1.26°F)
    0.9°C (1.62°F)
    Pt500 (385) -200.0 to 100.0°C (-328 to 212°F)
    100.0 to 400.0°C (212 to 752°F)
    400.0 to 630.0°C (752 to 1166°F)
    0.3°C (0.54°F)
    0.4°C (0.72°F)
    0.3°C (0.5°F)
    Pt1000 (385) -200.0 to 300.0°C (-328 to 572°F)
    300.0 to 630.0°C (752 to 1166°F)<
    0.2°C (0.36°F)
    0.3°C (0.5°F)
    Pt100 (3926) -200.0 to -80.0°C (-328 to -112°F)
    -80.0 to 300.0°C (-112 to 572°F)
    300.0 to 630.0°C (572 to 1166°F)
    630.0 to 800.0°C (-328 to 32°F)
    0.1°C (0.18°F)
    0.2°C (0.36°F)
    0.3°C (0.54°F)
    0.6°C (1.8°F)
    Pt100 (3916) -200.0 to -80.0°C (-328 to -112°F)
    -80.0 to 300.0°C (-112 to 572°F)
    300.0 to 630.0°C (572 to 1166°F)
    630.0 to 800.0°C (-328 to 32°F)
    0.1°C (0.18°F)
    0.2°C (0.36°F)
    0.3°C (0.54°F)
    0.6°C (1.8°F)
    Ni120 (672) -80.0 to 260.0°C (-112 to 500°F) 0.1°C (0.18°F)
    Cu10 -100.0 to 260.0°C (-148 to 500°F) 1.3°C (2.34°F)
    Cu50 -180.0 to 200.0°C (-292 to 392°F) 0.3°C (0.54°F)
    Cu100 -180.0 to 200.0°C (-292 to 392°F) 0.1°C (0.18°F)
    YSI400 15.00 to 50.00 (59 to 122°F) 0.1°C (0.18°F)
    *Read Accuracy is based on 4-wire input. For 3-wire input add ±0.05Ω assuming all three RTD leads are matched.
    MC-1010 DC Voltage Measurement/Source Range Table
      Range Accuracy (% of reading ± floor)
    Read 0.000V to 20.000V 0.015% ± 2mV
    Source 0.000V to 20.000V 0.015% ± 2mV
    Maximum current output in voltage ranges is 3mA with an output impedance of <= 1Ω.
    MC-1010 DC mA Measurement/Source Range Table
      Range Accuracy (% of reading ± floor)
    Read 0.000mA to 24.000mA 0.015% ± 2µA
    Read 0.000mA - 24.000mA 0.015% ± 2µA
    Source 0.000mA - 24.000mA 0.015% ± 2µA
    Maximum load on mA source is 1000Ω. Voltage input range on simulate mode 5V to 30V.
    MC-1010 Frequency Measurement/Source Range Table
      Range Accuracy (% of reading ± floor)
    Read 2.0CPM to 600.0CPM
    1.0Hz - 1000.0Hz
    1.00KHz - 10.00KHz
    0.05% ± 0.1CPM
    0.05% ± 0.1Hz
    0.05% ± 0.01KHz
    Source 2.0CPM to 600.0CPM
    1.0Hz - 1000.0Hz
    1.00KHz - 10.00KHz
    0.05%
    0.05%
    0.25%
    Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude is adjustable from 1V to 20V, and is a square wave with 50% duty cycle.For output frequency, a slight negative offset of approximately -0.1V is present to assure zero crossing.
    MC-1010 Resistance Measurement/Source Range Table
      Range Excitation Current Accuracy (% of reading ± floor)
    Read Ohms Low 0.00Ω - 400.0Ω NA 0.025% ± 0.05Ω
    Rrad Ohms High 401.0Ω - 4000.0Ω NA 0.025% ± 0.5Ω
    Source Ohms Low 5.0Ω to 400.0Ω
    5.0Ω to 400.0Ω
    0.1mA to 0.5mA
    0.5mA to 3mA
    0.025% ± 0.1Ω
    0.025% ± 0.05Ω
    Source Ohms High 400Ω to 1500Ω
    1500Ω to 4000Ω
    0.05mA to 0.8mA
    0.05mA to 0.4mA
    0.025% ± 0.5Ω
    0.025% ± 0.05Ω
    Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.