MicroCal 200/200+ 2-Channel Multifunction Calibrator

|
| MicroCal 200+ Shown with Expansion PC Card Slot |
|
. Simultaneous Indication
of Input and Output with Engineering Units
. Large Easy to Read LCD Display
. Large Easy to Use Keys
. Change Parameter Fast Using
Front Keypad
. Isolated Input Channel 2 and
Output Channel 1
. Real Time Graphing and Data
Logging Capability

|
|
 |
Thermocouple Range and Accuracy Table
|
Thermocouple |
|
|
Microcal 200 |
Microcal 200+ |
|
Type |
Range |
Resolution |
Accuracy* |
Accuracy* |
|
J |
-210 to 1200°C |
0.1°C |
±(0.02% +0.1°C) |
±(0.01% +0.1°C) |
|
|
-346 to 2192°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
K |
-270 to 1370°C |
0.1°C |
±(0.02% +0.1°C) |
±(0.01% +0.1°C) |
|
|
-454 to 2498°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
T |
-270 to 400°C |
0.01°C |
±(0.02% +0.1°C) |
±(0.01% +0.1°C) |
|
|
-454 to 752°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
E |
-270 to 1000°C |
0.1°C |
±(0.02% +0.1°C |
±(0.01% +0.1°C) |
|
|
-454 to 1832°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
R/S |
-50 to 1760°C |
0.1°C |
±(0.02% +0.2°C) |
±(0.01% +0.2°C) |
|
|
-58 to 3200°F |
0.1°F |
±(0.02% +0.36°F) |
±(0.01% +0.36°F) |
|
B |
50 to 1820°C |
0.1°C |
±(0.02% +0.3°C) |
±(0.01% +0.3°C) |
|
|
122 to 3308°F |
0.1°F |
±(0.02% +0.54°F) |
±(0.01% +0.54°F) |
|
C |
0 to 2300°C |
0.1°C |
±(0.02% +0.2°C) |
±(0.01% +0.2°C) |
|
|
32 to 4172°F |
0.1°F |
±(0.02% +0.36°F) |
±(0.01% +0.36°F) |
|
G/D |
0 to 2300°C |
0.1°C |
±(0.02% +0.3°C) |
±(0.01% +0.3°C) |
|
|
32 to 4172°F |
0.1°F |
±(0.02% +0.54°F) |
±(0.01% +0.54°F) |
|
U |
-200 to 400°C |
0.1°C |
±(0.02% +0.1°C) |
±(0.01% +0.1°C) |
|
(J DIN) |
-328 to 752°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
L |
-200 to 760°C |
0.1°C |
±(0.02% +0.1°C) |
±(0.01% +0.1°C) |
|
(T DIN) |
-328 to 1400°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
N |
-270 to 1300°C |
0.1°C |
±(0.02% + 0.1°C) |
±(0.01% +0.1°C) |
|
|
-454 to 2372°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
F |
0 to 1400°C |
0.1°C |
±(0.02% +0.1°C) |
±(0.01% +0.1°C) |
|
(Platinel II) |
32 to 2552°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
RTD Range and Accuracy Table
|
RTD |
|
|
Microcal 200 |
Microcal 200+ |
|
Type |
Range |
Resolution |
Accuracy* |
Accuracy* |
|
Pt100 |
-200 to 850°C |
0.01°C |
±(0.02% +0.05°C) |
±(0.01% +0.05°C) |
|
385, 391, 3926 |
-328 to 1562 °F |
0.1°F |
±(0.02% +0.09°F) |
±(0.01% +0.09°F) |
|
Pt100 |
-200 to 650°C |
0.01°C |
±(0.02% +0.05°C) |
±(0.01% +0.05°C) |
|
3902 |
-328 to 1202°F |
0.1°F |
±(0.02% +0.09°F) |
±(0.01% +0.09°F) |
|
Pt100 |
-200 to 600°C |
0.01°C |
±(0.02% +0.05°C) |
±(0.01% +0.05°C) |
|
3916, 3923 |
-328 to 1112°F |
0.1°F |
±(0.02% +0.09°F) |
±(0.01% +0.09°F) |
|
Pt200 |
-200 to 850°C |
0.1°C |
±(0.02% +0.15°C) |
±(0.01% +0.15°C) |
|
|
-328 to 1562°F |
0.1°F |
±(0.02% +0.27°F) |
±(0.01% +0.27°F) |
|
Pt500 |
-200 to 850°C |
0.1°C |
±(0.02% +0.1°C) |
±(0.01% +0.1°C) |
|
|
-328 to
1562°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
Pt1000 |
-200 to 850°C |
0.01°C |
±(0.02% +0.1°C) |
±(0.01% +0.1°C) |
|
385, 391 |
-328 to 1562°F |
0.1°F |
±(0.02% +0.18°F) |
±(0.01% +0.18°F) |
|
Cu10 |
-70 to 150°C |
0.1°C |
±(0.02% +0.4°C) |
±(0.01% +0.4°C) |
|
|
-94 to 302°F |
0.1°F |
±(0.02% +0.72°F) |
±(0.01% +0.72°F) |
|
Cu100 |
-180 to 150°C |
0.1°C |
±(0.02% +0.05°C) |
±(0.01% +0.05°C) |
|
|
-292 to
302°F |
0.1°F |
±(0.02% +0.09°F) |
±(0.01% +0.09°F) |
|
Ni100 |
-60 to 180°C |
0.1°C |
±(0.02% +0.05°C) |
±(0.01% +0.05°C) |
|
|
-76 to 356°F |
0.1°F |
±(0.02% +0.09°F) |
±(0.01% +0.09°F) |
|
Ni120 |
0 to 150°C |
0.1°C |
±(0.02% +0.05°C) |
±(0.01% +0.05°C) |
|
|
32 to 302°F |
0.1°F |
±(0.02% +0.09°F) |
±(0.01% +0.09°F) |
Process Signals Range and Accuracy Table
|
Process |
|
|
Microcal 200 |
Microcal 200+ |
|
Signal Type |
Range |
Resol. |
Accuracy* |
Accuracy* |
|
Ohms IN |
0 to 500 Ohms |
1 mOhms |
±(0.01% +12mOhms) |
±(0.02% +12mOhms) |
|
|
0 to 5000 Ohms |
0.01 Ohms |
±(0.01% +120mOhms) |
±(0.02% +120mOhms) |
|
Ohms OUT |
0 to 500 Ohms |
1 mOhms |
±(0.01% +20mOhms) |
±(0.02% +20mOhms) |
|
|
0 to 5000 Ohms |
0.01 Ohms |
±(0.01% +200mOhms) |
±(0.02% +200mOhms) |
|
mA (In) |
-5 to +50mA |
0.1µA |
±(0.01% +0.4µA) |
±(0.02% +0.4µA) |
|
mA (Out) |
0 to +50mA |
0.1µA |
±(0.01% +0.4µA) |
±(0.02% +0.4µA |
|
mV (L) |
-20 to +200mV |
1µV |
±(0.01% +2µV) |
±(0.02% +2µV) |
|
mV (H) |
-0.2 to +2 V |
10µV |
±(0.01% +10µV |
±(0.02% +10µV) |
|
V |
-2 to +20 V |
0.1mV |
±(0.01% +0.08 mV) |
±(0.02% +0.08mV) |
|
|
1 to 200 Hz |
0.001 Hz |
±(0.005% +0.001 Hz) |
|
|
Frequency |
1 to 2,000 Hz |
0.01 |
±(0.005% +0.001 Hz) |
|
|
|
1 to 20,000 Hz |
0.1 Hz |
±(0.005% +0.001 Hz) |
|
|
Pulse counter |
0 to 106 counts |
1 count |
infinite |
infinite |
|
Pulse (Out) |
0 to 6,000 pulse/min |
1 pulse/min |
1 pulse/min |
1 pulse/min |
|
|
0 to 36,000 pulse/h |
1 pulse/hr |
1 pulse/min |
1 pulse/min |
Note: The MicroCal 200/200+ relative accuracies shown above are stated for 360 days and the operative conditions are from +64°F to +82°F. Typical 90 days relative accuracy can be estimated by dividing the "% of reading" specifications by 1.6. Typical
2 years relative accuracy can be estimated by multiplying the “% of
reading" specifications by 1.4. All input ranges:
additional error ±1 digit.

|
| MicroCal 200+ Shown |
|
Click Here To Order: Model Microcal 200 or 200+ Multifunction Calibrator. Please
call for Price and Availability
Each instrument includes USB Serial Interface, Real time clock and internal memory, Soft Vinyl protection
case, NIST Certificate of Calibration, Instruction Manual, Spare fuses and Logman, LinMan, Calpman PC software.
|
|
 |
|
|
|
|
|
Multifunction Temperature and Process Calibrators, Thermocouple and mA Calibrators, Thermocouple and RTD Dry Block Calibrators, Portable Infrared Thermometers, Fixed Mounted Industrial Infrared Thermometers, Thermocouple and RTD 2-wire Head Mounted Transmitters, DIN Rail Mounted Thermocouple and RTD Transmitters, Temperature and Process Panel Meters, Temperature & Humidity Transmitters, Hand Held Digital Thermocouple Thermometers, MIST Microprocessor Based Integral Transmitter RTD Temperature
Sensors,
Industrial Test tools and Maintenance Kit, Indoor Air
Quality (IAQ) and Test Meters, Hand Held Humidity and Air Flow Meters, Industrial Portable
Refrigerant and Gas Leak Detection Meters, Pressure Measurement and Calibration Instruments
Temperature & Process Instruments Inc. * 1767 Central Park Avenue * Suite 112 * Yonkers