    The DMC-1410 Documenting Multifunction Calibrator provides a feature set unmatched in high accuracy, hand-held calibrators in its price range. The DMC-1410 provides the functions and accuracy associated with fixed installation, laboratory instruments, and has everything needed for virtually any calibration task. What makes this versatile calibrator best in class is the ability to fully document any calibration easily while performing the normal job. No special software is required! You already have all the software you need installed on your PC. Some functions will require a Microsoft® Excel®-compatible spreadsheet.
    Features
    • Measure and simulate Thermocouples (13 types), RTDs (13 types), Ohms, current, voltage, frequency and source pulse trains
    • Test for pass/fail criteria
    • Capture both AS FOUND and AS LEFT data
    • Review data and print calibration certificates in the field
    • Upload data to PC for analysis with Microsoft® Excel® or full database
    • Built-in 250 Ohm resistor for HART compatibility with smart transmitters and PLCs
    • Built-in 24 V supply can drive 4 to 20mA loops up to 1000 Ohms
    • Store up to 21 data points per instrument, and up to 50 instruments in non-volatile memory
    • Highest accuracy in class to 0.015% of reading
    • Direct entry of custom RTD coefficients (R0, A, B, C)
    • Isolated mA/V read-back circuit for complete transmitter calibration
      Shipping Please Call for availability
    914-673-0333

    DMC-1410 Documenting Temperature and Process Calibrator Features
    We call this “the documenting calibrator for the rest of us.” Work better - get better work. Set up and calibrate any instrument as you normally would. Just press the “SAVE” key after each calibration point. When you complete the calibration, follow the prompts to enter the “tag” information and you are DONE. The DMC-1410 also features the ability to test for pass/fail criteria, capture both AS FOUND and AS LEFT data, maintain calibration data for 50 instruments in nonvolatile memory, and even print calibration certificates right in the field, with no PC needed, using our optional OTS-100 portable certificate printer. Measure and simulate thermocouples, RTDs, current, voltage, frequency and source pulse trains with one lightweight easy to use calibrator.
    HART Compatibility
    A built-in 250 Ohm resistor is provided for HART™ compatibility with smart transmitters and PLCs. Full fuseless protection and a serial communications port for full control with ASCII commands, are just some of the additional features that make the DMC-1400 the single, most indispensable tool available for virtually any calibration task.
    DMC-1410 Temperature and Process Calibrator General Specifications
    ClearBriteTM Display High contrast 128 x 64 pixel addressable graphic LCD w/daylight backlight 2.4 x 1.8 in., 63 x 44 mm
    Serial Communications Yes, ASCII, RS-232, requires optional Martel 1919069 serial cable or 1919896 USB cable
    Operating Temperature -10°C to +50°C (-14°F to 122°F)
    Storage Temperature -20°C to +70°C (-4°F to 158°F)
    Power Requirements 6Vdc; Batteries 4 AA alkaline (supplied) or optional rechargeable
    CE - EMC EN50082-1 and EN55022: 1994 Class B
    Safety CSA C22.2 No. 1010-1: 1992
    Dimensions 8.5" H x 4.8" W x 2.1" D (22 x 12 x 5.3 cm)
    Weight 1.8 lbs. (0.82 kg)
    Other IP54
    Protected against misconnection to 250 VAC/VDC
    DMC-1410 Thermocouple Range and Accuracy Table
    Sensor Type Temperature Range Accuracy with CJC On
    Type J -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 800.0°C (32 to 1472°F)
    800.0 to 1200.0 (1472 to 2192°F)
    0.6°C (1.1°F)
    0.4°C (0.7°F)
    0.5°C (0.9°F)
    Type K -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 1000.0°C (32 to 1832°F)
    1000.0 to 1372.0°C (1832 to 2498°F)
    0.8°C (1.4°F)
    0.5°C (0.9°F)
    0.7°C (1.3°F)
    Type T -250.0 to 0.0°C (-418 to 32°F)
    0.0 to 400.0°C (32 to 752°F)
    0.8°C (1.4°F)
    0.5°C (0.9°F)
    Type E -250.0 to -100.0°C (-418 to -148°F)
    -100.0 to 1000.0°C (-148 to 1832°F)
    0.8°C (1.5°F)
    0.4°C (0.7°F)
    Type R 0 to 1767°C (32 to 3203°F) 1.4°C (2.5°F)
    Type S 0 to 1767°C (32 to 3203°F) 1.4°C (2.5°F)
    Type B 600 to 800°C (1112 to 1472°F)
    800 to 1000°C (1472 to 1832°F)
    1000 to 1800°C (1832 to 3272°F)
    1°C (1.8°F)
    1.5°C (2.7°F)
    1.7°C (3°F)
    Type C 0 to 1000°C (32 to 1832°F)
    1000 to 2318°C (1832 to 4204°F)
    0.8°C (1.4°F)
    2.5°C(4.5°F)
    Type L -200 to 0.0°C (-328 to 32°F)
    0.0 to 900.0°C (32 to 1652°F)
    0.5°C (0.9°F)
    0.4°C(0.7°F)
    Type U -200 to 0.0°C (-328 to 32°F)
    0.0 to 400.0°C (32 to 752°F)
    0.7°C (1.3°F)
    0.45°C (0.8°F)
    Type N -200 to 0.0°C (-328 to 32°F)
    -0 to 600.0°C (32 to 1112°F)
    600.0 to 1300.0°C (1112 to 2372°F)
    1°C (1.8°F)
    0.5°C (.9°F)
    0.6°C (1.1°F)
    mV Read and Source -10.000mV to 75.000mV 0.015% ± 10µV
    Specifications (All specifications apply at 23 °C ± 5°C unless otherwise stated. Outside this range add ±0.005% of reading/°C. Accuracy is % of reading ± floor spec.)
    DMC-1410 RTD Range and Accuracy Tables
    Sensor Type Temperature Range Accuracy
    Pt10 (385) -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 100.0°C (32 to 212°F)
    100.0 to 300.0°C (212 to 572°F)
    300.0 to 400.0°C (572 to 752°F)
    400.0 to 630.0°C (752 to 1166°F)
    630.0 to 800.0°C (1166 to 1472°F)
    0.78°C (1.4°F)
    0.8°C (1.5°F)
    0.9°C (1.8°F)
    0.98°C (1.9°F)
    1.05°C (1.9°F)
    1.16°C (2.0°F)
    Pt50 (385) -200.0 to 0.0°C (-328 to 32°F)
    0.0 to 100.0°C (32 to 212°F)
    100.0 to 300.0°C (212 to 572°F)
    300.0 to 400.0°C (572 to 752°F)
    400.0 to 630.0°C (752 to 1166°F)
    630.0 to 800.0°C (1166 to 1472°F)
    0.23°C (0.4°F)
    0.15°C (0.27°F)
    0.15°C (0.27°F)
    0.18°C (0.3°F)
    0.21°C (0.38°F)
    0.26°C (0.5°F)
    Pt100 (385) -200.0 to -80.0°C (-328 to -112°F)
    -80.0 to 300.0°C (-112 to 572°F)
    300.0 to 630.0°C (572 to 1166°F)
    630.0 to 800.0°C (-328 to 32°F)
    0.08°C (0.14°F)
    0.15°C (0.26°F)
    0.21°C (0.38°F)
    >0.26°C (0.5°F)
    Pt200 (385) -200.0 400.0°C (-328 to 752°F)
    400.0 to 630.0°C (752 to 1166°F)
    0.4°C (0.72°F)
    0.5°C (0.9°F)
    Pt500 (385) -200.0 to 100.0°C (-328 to 212°F)
    100.0 to 400.0°C (212 to 752°F)
    400.0 to 630.0°C (752 to 1166°F)
    0.19°C (0.34°F)
    0.25°C (0.45°F)
    0.52°C (0.9°F)
    Pt1000 (385) -200.0 to 300.0°C (-328 to 572°F)
    300.0 to 630.0°C (752 to 1166°F)<
    0.14°C (0.25°F)
    0.19°C (0.34°F)
    Pt100 (3926) -200.0 to -80.0°C (-328 to -112°F)
    -80.0 to 300.0°C (-112 to 572°F)
    300.0 to 630.0°C (572 to 1166°F)
    0.07°C (0.12°F)
    0.13°C (0.23°F)
    0.19°C (.34°F)
    Pt100 (3916) -200.0 to -80.0°C (-328 to -112°F)
    -80.0 to 300.0°C (-112 to 572°F)
    300.0 to 630.0°C (572 to 1166°F)
    0.07°C (0.12°F)
    0.17°C (0.3°F)
    0.19°C (0.34°F)
    Ni120 (672) -80.0 to 260.0°C (-112 to 500°F) 0.1°C (0.18°F)
    Cu10 -100.0 to 260.0°C (-148 to 500°F) 1.3°C (2.34°F)
    Cu50 -180.0 to 200.0°C (-292 to 392°F) 0.3°C (0.54°F)
    Cu100 -180.0 to 200.0°C (-292 to 392°F) 0.1°C (0.18°F)
    YSI400 15.00 to 50.00 (59 to 122°F) 0.1°C (0.18°F)
    *Read Accuracy is based on 4-wire input. For 3-wire input add ±0.05Ω assuming all three RTD leads are matched.
    DMC-1410 DC Voltage Measurement/Source Range Table
      Range Accuracy (% of reading ± floor)
    Read Isolated
    (Upper Display)
    0.000V to 30.000V 0.015% ± 2mV
    Read Non-Isolated
    (Lower Display)
    0.000V to 20.000V 0.015% ± 2mV
    Source 0.000V to 20.000V 0.015% ± 2mV
    Maximum current output in voltage ranges is 3mA with an output impedance of <= 1Ω.
    DMC-1410 DC mA Measurement/Source Range Table
      Range Accuracy (% of reading ± floor)
    Read Isolated
    (Upper Display)
    0.000mA to 24.000mA 0.015% ± 2µA
    Read Non-Isolated
    (Lower Display)
    0.000mA - 24.000mA 0.015% ± 2µA
    Source 0.000mA - 24.000mA 0.015% ± 2µA
    Maximum load on mA source is 1000Ω. Voltage input range on simulate mode 5V to 30V.
    DMC-1410 Frequency Measurement/Source Range Table
      Range Accuracy (% of reading ± floor)
    Read 2.0CPM to 600.0CPM
    1.0Hz - 1000.0Hz
    1.00KHz - 10.00KHz
    0.05% ± 0.1CPM
    0.05% ± 0.1Hz
    0.05% ± 0.01KHz
    Source 2.0CPM to 600.0CPM
    1.0Hz - 1000.0Hz
    1.00KHz - 10.00KHz
    0.05%
    0.05%
    0.25%
    Input voltage amplitude range on frequency is 1V to 20V zero based square wave only. Output amplitude is adjustable from 1V to 20V, and is a square wave with 50% duty cycle. For output frequency, a slight negative offset of approximately -0.1V is present to assure zero crossing.
    DMC-1410 Resistance Measurement/Source Range Table
      Range Excitation Current Accuracy (% of reading ± floor)
    Read Ohms Low 0.00Ω - 400.0Ω NA 0.025% ± 0.05Ω
    Read Ohms High 401.0Ω - 4000.0Ω NA 0.025% ± 0.5Ω
    Source Ohms Low 5.0Ω to 400.0Ω
    5.0Ω to 400.0Ω
    0.1mA to 0.5mA
    0.5mA to 3mA
    0.025% ± 0.1Ω
    0.025% ± 0.05Ω
    Source Ohms High 400Ω to 1500Ω
    1500Ω to 4000Ω
    0.05mA to 0.8mA
    0.05mA to 0.4mA
    0.025% ± 0.5Ω
    0.025% ± 0.05Ω
    Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.