DMC-1410 High Accuracy Multifunction Temperature and Process Documenting Calibrator with NIST Calibration Certificate
Special Price $2754.00 save 5%
New
Precision
Multifunction Calibrator with ClearBrite™ High Graphic Dual Display
|
|
DMC-1410 Multifunction Process Calibrator |

|
| Supplied with Protective Rubber Boot, Test Leads, USB Adaptor and NIST Cetrificate |
Prices Start at $2899.00
Special Introductory Price $2754.00 save 5%

|
Environmental Specifications
Operating Temperature: -10°C to +50°C (-14°F to 122°F) Storage
Temperature: -20°C to +70°C (-4°F to 158°F) Stability: ±0.005% of reading/°C
outside of 23 °C ± 5°C Power Requirements: 6Vdc; Batteries 4 AA alkaline (supplied) or optional rechargeable
Mechanical Specifications
Dimensions:
8.5" H x 4.8" W x 2.1"
D
(22 x 12 x 5.3
cm)
Weight:
1.8 lbs (0.82 kg)
Signal Specifications
Voltage
Read and Source Source: 0.000 to 20.000 VDC
Read: Isolated 0.000 to 30.000 VDC Non-Isolated 0.000
to 20.000 VDC Thermocouple Types and Ranges: See
table below Accuracy: See Table Below
Temperature scale:
ITS90
mV
Read/Source:
-10.000 to +75.000 mV
Current (mA) Source:
0.000 to 24.000mA Read: Isolated: 0.000 to 24.000mA; Non-Isolated: 0.000 to 24.000mA Frequency (1 to 20 V selectable amplitude)
CPM Source and Read: 2.0
to 600.0 CPM Hz Source and Read: 1.0 to 1000.0 Hz
kHz Source and Read: 1.0 to 10.00 kHz Pulse (Source only; 1 to 20 V selectable amplitude)
Pulses: 1 to 30,000.0 - 2 CPM to
10 kHz Ohms
Source: 5.0 to 4000 Ohms
Read: 0.00 to 4000.0
Ohms
|
Thermocouple Range and Accuracy
| TC Type | Range °C (°F) | Accuracy* | | | -210.0 to 0.0°C (-346 to
32°F) | 0.4°C
(0.7°F) |
| J | 0 to
800.0°C (32 to 1472°F) | 0.2°C (0.4°F) |
| | 800.0 to 1200.0°C (1472 to 2192°F) | 0.3°C (0.5°F) |
| | -200.0 to 0.0°C (-328 to 32°F) | 0.6°C (1.1°F) |
| K | 0 to 1000°C (32 to 1832°F) | 0.3°C (0.5°F) |
| | 1000.0 to 1372.0°C (1832 to 2502°F) | 0.5°C (0.9°F) |
| T | -250.0 to 0.0°C (-418 to 32°F) | 0.6°C (1.1°F) |
| | 0 to 400.0°C (32 to 752°F) | 0.2°C (0.4°F) |
| E | -250.0 to -100.0°C (-418 to 212°F) | 0.6°C (1.1°F) |
| | -100.0 to 1000.0°C (-148 to 1832°F) | 0.2°C (0.4°F) |
| R | 0 to 1767.0°C (32 to 3212°F) | 1.2°C (2.2°F) |
| S | 0 to 1767.0°C (32 to 3212°F) | 1.2°C (2.2°F) |
| | 600.0 to 800.0°C (1112 to 1472°F) | 1.2°C (2.2°F) |
| B | 800.0 to 1000.0°C (1472 to 18322°F) | 1.3°C (2.4°F) |
| | 1000.0 to 1820.0°C (1832 to 3308°F) | 1.5°C (2.7°F) |
| C | 0 to 1000.0°C (32 to 1832°F) | 0.6°C (1.1°F) |
| | 1000.0 to 2316.0°C (1832 to 3320°F) | 2.3°C (4.2°F) |
| XK | -200.0 to 800.0°C (-328 to 1472°F) | 0.2°C (0.4°F) |
| BP | 0 to 800.0°C (32 to 1472°F) | 0.9°C (1.6°F) |
| | 800.0 to 2500.0°C (1472 to 4532°F) | 2.3°C (4.2°F) |
| L | -200.0 to 0.0°C (-328 to 32°F) | 0.25°C (0.45°F) |
| | 0 to 900.0°C (32 to 1652°F) | 0.2°C (0.4°F) |
| U | -200.0 to 0.0°C (-328 to 32°F) | 0.5°C (0.9°F) |
| | 0 to 600.0°C (32 to 1112°F) | 0.25°C (0.45°F) |
| N | -200.0 to 0.0°C (-328 to 32°F) | 0.8°C (1.4°F) |
| | 0 to 1300.0°C (32 to 2370°F) | 0.4°C (0.7°F) |
*All TC errors include CJC errors
CJC error outside of 23 ±
5°C is 0.05°C/°C (In °C add 0.2 for cold junction
compensation error.)
|
RTD Type |
Range |
Accuracy* |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.2°C (0.4°F) |
|
Pt100 (385) |
100.0 to 300.0°C (212 to 572°F) |
0.3°C (0.5°F) |
|
|
300.0 to 600.0°C (572 to 1112°F) |
0.4°C (0.7°F) |
|
|
600.0 to 800.0°C (1112 to 1472°F) |
0.5°C (0.9°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.8°C (1.4°F) |
|
Pt200 (385) |
100.0 to 300.0°C (212 to 572°F) |
0.9°C (1.9°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
1.0°C (1.8°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.4°C (0.7°F) |
|
Pt500 (385) |
100.0 to 300.0°C (212 to 572°F) |
0.5°C (0.9°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
0.6°C (1.1°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.2°C (0.4°F) |
|
Pt1000 (385) |
100.0 to 300.0°C (212 to 572°F) |
0.3°C (0.5°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
0.4°C (0.7°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
1.4°C (2.5°F) |
|
Pt385-10 |
100.0 to 300.0°C (212 to 572°F) |
1.6°C (2.9°F) |
|
|
300.0 to 600.0°C (572 to 1112°F) |
1.8°C (3.2°F) |
|
|
600.0 to 800.0°C (1112 to 1472°F) |
2.0°C (3.6°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.4°C (0.7°F) |
|
Pt385-50 |
100.0 to 300.0°C (212 to 572°F) |
0.5°C (0.9°F) |
|
|
300.0 to 600.0°C (572 to 1112°F) |
0.6°C (1.1°F) |
|
|
600.0 to 800.0°C (1112 to 1472°F) |
0.7°C (1.3°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.2°C (0.4°F) |
|
Pt100 (3926) |
100.0 to 300.0°C (212 to 572°F) |
0.3°C (0.5°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
0.4°C (0.7°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.2°C (0.4°F) |
|
Pt100 (3916) |
100.0 to 300.0°C (212 to 572°F) |
0.3°C (0.5°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
0.4°C (0.7°F) |
|
Ni120 (672) |
-80.0 to 260.0°C (-112 to 500°F) |
0.2°C (0.4°F) |
|
Cu10 |
-100.0 to 260.0°C (-148 to 500°F) |
1.4°C (2.5°F) |
|
Cu50 |
-180.0 to 200.0°C (-292 to 392°F) |
0.4°C (0.7°F) |
|
Cu100 |
-180.0 to 200.0°C (-292 to 392°F) |
0.3°C (0.5°F) |
|
YSI400 |
15.00 to 50.00 (59 to 122°F) |
0.1°C (0.2°F) |
*Read
Accuracy is based on 4-wire input. For 3-wire input add ±0.05W
assuming all three RTD leads are matched.
| Process Signals | Read Accuracy* | Source Accuracy* |
|
Current (mA) | ±0.015%
of reading,±2 μA | ±0.015% of reading,
±2 μA |
|
CPM | ±0.05% of reading,±1 LSD | ±0.05% of reading | |
Hz | ±0.05% of reading,±1 LSD | ±0.05%
of reading |
|
kHz | ±0.05% of reading, ±1 LSD | ±0.125% of reading | | Ohms works with
all pulsed transmitters down to 5 ms | | |
|
400 Ohm Range | ±0.025% of reading, ±0.05W | ±0.025% of reading, ±0.05W |
|
4000 Ohm Range | ±0.025% of reading, ±0.5W
| ±0.025% of reading, ±0.5W |
*DMC-1400 Accuracy Specifications (23 °C ±5 °C unless otherwise noted)
|
| |
|
|
|
The DMC-1410 is supplied with a
rugged protective boot, four
standard alkaline AA batteries,
padded carrying case, test
leads, communications cable,
USB adapter, user manual, quick
start guide, DVD and CD based
video training and, NIST
traceable calibration certificate.
|
|
 |
|
|
|
Multifunction Temperature and Process Calibrators,
Thermocouple and mA Calibrators,
Thermocouple and RTD Dry Block Calibrators,
Portable Infrared Thermometers,
Fixed Mounted Industrial Infrared Thermometers,
Thermocouple and RTD 2-wire Head Mounted Transmitters, DIN Rail Mounted Thermocouple and RTD Transmitters,
Temperature and Process Panel Meters, Temperature & Humidity Transmitters, Hand Held Digital Thermocouple Thermometers,
MIST Microprocessor Based Integral Transmitter RTD Temperature
Sensors,
Industrial Test tools and Multimeters, Indoor Air
Quality (IAQ) and Test Meters,
Hand Held Humidity and Air Flow Meters,
Industrial Portable
Refrigerant and Gas Leak Detection Meters,
Pressure Measurement and Calibration Instruments,
Boiler Flue Gas
Combustion Emission Analyzers,
Solar Radiation Meter
Temperature & Process Instruments Inc. * 1767 Central Park Avenue * Suite 112 * Yonkers