Welcome to Temperature & Process Instruments, Inc.

DMC1400 Multifunction Documenting Calibrator Specifications

Home
Shop Online
Featured Products and Specials
Request a Quotation or Product Information
Microprocessor Integrated Sensor Transmitter (MIST) Temperature Sensors
Portable Non-Contact Infrared Thermometers
Industrial Fixed Mount Infrared Thermometers
Multifunction Temperature & Process Calibrators
Single Function Temperature & Process Calibrators
Dry Block Thermocouple & RTD Calibrators
Head Mounted Programmable Temperature Transmitters
DIN Rail Mounted Programmable Temperature Transmitters
Temperature & Process Panel Meters
Temperature & Humidity Transmitters
Hand Held Indoor Air Quality (IAQ) and Test Instruments
Hand Held Temperature, Humidity, and Air Flow Instruments
Industrial Test Tools
Portable Thermocouple Thermometers/Dataloggers
Industrial Refrigerant and Gas Leak Detection
Pressure Measurement and Calibration Instruments
New Products Coming Soon
Technical Information
Application Notes
Documentation and User Manuals
About Us
Shipping and Policies
Contact Us
Website Map

DMC 1400 High Accuracy Multifunction Temperature and Process Documenting Calibrator with NIST Calibration Certificate

Precision Dual Display Multifunction Calibrator
DMC 1400 Multifunction Process Calibrator
DMC1400-whatsincluded-small.jpg
Supplied with Protective Rubber Boot, Test Leads USB Adaptor and NIST Cetrificate
Prices Start at $2695.00

Click here to Order MC1200 Multifunction Temperature Pressure and Process Calibrator.gif

Environmental Specifications
Operating Temperature: -10°C to +50°C (-14°F to 122°F)
Storage Temperature: -20°C to +70°C (-4°F to 158°F)
Stability: ±0.005% of reading/°C outside of 23 °C ± 5°C
Power Requirements: 6Vdc; Batteries 4 AA alkaline (supplied) or optional rechargeable
Mechanical Specifications
Dimensions:
8.7" H x 4.2" W x 2.3" D (220.9 x 106.6 x 58.4 mm)
Weight: 30.5 ounces (863 grams)

Signal Specifications
Voltage Read and Source
Source:
0.000 to 20.000 VDC
Read: Isolated 0.000 to 30.000 VDC
Non-Isolated 0.000 to 20.000 VDC
Thermocouple
Types and Ranges: See table below
Accuracy: See Table Below
Temperature scale: ITS90
mV
Read/Source: -10.000 to +75.000 mV
Current (mA)
Source:
0.000 to 24.000mA
Read: Isolated: 0.000 to 24.000mA; 
Non-Isolated: 0.000 to 24.000mA
Frequency (1 to 20 V selectable amplitude)
CPM Source and Read: 2.0 to 600.0 CPM
Hz Source and Read: 1.0 to 1000.0 Hz
kHz Source and Read: 1.0 to 10.00 kHz
Pulse (Source only; 1 to 20 V selectable amplitude) 
Pulses: 1 to 30,000.0 - 2 CPM to 10 kHz
Ohms
Source: 5.0 to 4000 Ohms
Read: 0.00 to 4000.0 Ohms

Thermocouple Range and Accuracy
TC Type Range °C (°F) Accuracy*
 -210.0 to  0.0°C (-346 to 32°F)0.4°C (0.7°F)
J0  to 800.0°C (32 to 1472°F)0.2°C (0.4°F)
 800.0 to 1200.0°C (1472 to 2192°F)0.3°C (0.5°F)
 -200.0 to 0.0°C (-328 to 32°F)0.6°C (1.1°F)
K0 to 1000°C (32 to 1832°F)0.3°C (0.5°F)
 1000.0 to 1372.0°C (1832 to 2502°F)0.5°C (0.9°F)
T-250.0 to 0.0°C (-418 to 32°F)0.6°C (1.1°F)
 0 to 400.0°C (32 to 752°F)0.2°C (0.4°F)
E-250.0 to -100.0°C (-418 to 212°F)0.6°C (1.1°F)
 -100.0 to 1000.0°C (-148 to 1832°F)0.2°C (0.4°F)
R0 to 1767.0°C (32 to 3212°F)1.2°C (2.2°F)
S0 to 1767.0°C (32 to 3212°F)1.2°C (2.2°F)
 600.0 to 800.0°C (1112 to 1472°F)1.2°C (2.2°F)
B800.0 to 1000.0°C (1472 to 18322°F)1.3°C (2.4°F)
 1000.0 to 1820.0°C (1832 to 3308°F)1.5°C (2.7°F)
C0 to 1000.0°C (32 to 1832°F)0.6°C (1.1°F)
 1000.0 to 2316.0°C (1832 to 3320°F)2.3°C (4.2°F)
XK-200.0 to 800.0°C (-328 to 1472°F)0.2°C (0.4°F)
BP0 to 800.0°C (32 to 1472°F)0.9°C (1.6°F)
 800.0 to 2500.0°C (1472 to 4532°F)2.3°C (4.2°F)
L-200.0 to 0.0°C (-328 to 32°F)0.25°C (0.45°F)
 0 to 900.0°C (32 to 1652°F)0.2°C (0.4°F)
U-200.0 to 0.0°C (-328 to 32°F)0.5°C (0.9°F)
 0 to 600.0°C (32 to 1112°F)0.25°C (0.45°F)
N-200.0 to 0.0°C (-328 to 32°F)0.8°C (1.4°F)
 0 to 1300.0°C (32 to 2370°F)0.4°C (0.7°F)
*All TC errors include CJC errors CJC error outside of 23 ± 5°C is 0.05°C/°C
(In °C add 0.2 for cold junction compensation error.)
 

RTD Type

Range

Accuracy*

 

-200.0 to 100.0°C (-328 to 212°F)

0.2°C (0.4°F)

Pt100 (385)

100.0 to 300.0°C (212 to 572°F)

0.3°C (0.5°F)

 

300.0 to 600.0°C (572 to 1112°F)

0.4°C (0.7°F)

 

600.0 to 800.0°C (1112 to 1472°F)

0.5°C (0.9°F)

 

-200.0 to 100.0°C (-328 to 212°F)

0.8°C (1.4°F)

Pt200 (385)

100.0 to 300.0°C (212 to 572°F)

0.9°C (1.9°F)

 

300.0 to 630.0°C (572 to 1166°F)

1.0°C (1.8°F)

 

-200.0 to 100.0°C (-328 to 212°F)

0.4°C (0.7°F)

Pt500 (385)

100.0 to 300.0°C (212 to 572°F)

0.5°C (0.9°F)

 

300.0 to 630.0°C (572 to 1166°F)

0.6°C (1.1°F)

 

-200.0 to 100.0°C (-328 to 212°F)

0.2°C (0.4°F)

Pt1000 (385)

100.0 to 300.0°C (212 to 572°F)

0.3°C (0.5°F)

 

300.0 to 630.0°C (572 to 1166°F)

0.4°C (0.7°F)

 

-200.0 to 100.0°C (-328 to 212°F)

1.4°C (2.5°F)

Pt385-10

100.0 to 300.0°C (212 to 572°F)

1.6°C (2.9°F)

 

300.0 to 600.0°C (572 to 1112°F)

1.8°C (3.2°F)

 

600.0 to 800.0°C (1112 to 1472°F)

2.0°C (3.6°F)

 

-200.0 to 100.0°C (-328 to 212°F)

0.4°C (0.7°F)

Pt385-50

100.0 to 300.0°C (212 to 572°F)

0.5°C (0.9°F)

 

300.0 to 600.0°C (572 to 1112°F)

0.6°C (1.1°F)

 

600.0 to 800.0°C (1112 to 1472°F)

0.7°C (1.3°F)

 

-200.0 to 100.0°C (-328 to 212°F)

0.2°C (0.4°F)

Pt100 (3926)

100.0 to 300.0°C (212 to 572°F)

0.3°C (0.5°F)

 

300.0 to 630.0°C (572 to 1166°F)

0.4°C (0.7°F)

 

-200.0 to 100.0°C (-328 to 212°F)

0.2°C (0.4°F)

Pt100 (3916)

100.0 to 300.0°C (212 to 572°F)

0.3°C (0.5°F)

 

300.0 to 630.0°C (572 to 1166°F)

0.4°C (0.7°F)

Ni120 (672)

-80.0 to 260.0°C (-112 to 500°F)

0.2°C (0.4°F)

Cu10

-100.0 to 260.0°C (-148 to 500°F)

1.4°C (2.5°F)

Cu50

-180.0 to 200.0°C (-292 to 392°F)

0.4°C (0.7°F)

Cu100

-180.0 to 200.0°C (-292 to 392°F)

0.3°C (0.5°F)

YSI400

15.00 to 50.00 (59 to 122°F)

0.1°C (0.2°F)

*Read Accuracy is based on 4-wire input. For 3-wire input add ±0.05W
assuming all three RTD leads are matched.

Process Signals

Read Accuracy*

Source Accuracy*

Current (mA) ±0.015% of reading,±2 μA±0.015% of reading, ±2 μA
CPM±0.05% of reading,±1 LSD ±0.05% of reading
Hz±0.05% of reading,±1 LSD±0.05% of reading
kHz ±0.05% of reading, ±1 LSD ±0.125% of reading
Ohms works with all pulsed transmitters down to 5 ms  
400 Ohm Range ±0.025% of reading, ±0.05W±0.025% of reading, ±0.05W
4000 Ohm Range ±0.025% of reading, ±0.5W ±0.025% of reading, ±0.5W

*DMC-1400 Accuracy Specifications (23 °C ±5 °C unless otherwise noted)

DMC1400-whatsincluded-small.jpg

The DMC-1400 is supplied with a rugged protective boot, four standard alkaline AA batteries, padded carrying case, test leads, communications cable, USB adapter, user manual, quick start guide, DVD and CD based video training and, of course, an NIST traceable calibration certificate.

pdficon.gif

Click Here to Download the DMC-1400 Specification Sheet


Temperature & Process Instruments Inc. * 1767 Central Park Avenue * Suite 112 * Yonkers  * NY * 10710
Phone: 1-914-673-0333 Fax: 1-866-292-1456 Email: