DC80R Simulate and Measure RTD Temperature and Ohm Signals Process Calibrator

DC80R RTD Temperature and Ohm Calibrator
    Price Starting At: $780.00

    Overview

    The DC80R RTD Process Calibrator measures and simulates resistance values of 7 distinct types of RTDs in °C or °F and Ohms. The DC80R RTD calibrator is ideal to either measure or simulate RTD Temperature Signals or Ohms to meet your test and calibration needs in the field or laboratory. It is an easy to use, lightweight and yet robust and are invaluable tools for the everyday use in plant instrumentation.
    Features
    • Measures and simulates SEVEN types of RTDs: Pt10, Pt50, Pt100 (385), Pt100 (392), Pt200, Pt500 and Pt1000
    • Generates and measures resistance values from 0 to 3200Ω
    • Resolution: RTD 0.1°C/°F; Resistance 0.01/0.1Ω
    • Accuracy of ±0.2°C for temperature
    • Accuracy of 0.1Ω for resistance
    • Temperature unit selection from °C and °F
    • Low battery indication
      Shipping Estimated Shipping 1 to 3 Weeks

    DC80-R RTD Temperature Calibrator Specifications
    Operating Temperature -10°C to +55°C (14°F to 131°F)
    Storage Temperature -40°C to +60°C (-40°F to 140°F)
    Temperature coefficient ±0.01%/°C (0°C to 18°C and 28°C to 50°C)
    Resolution RTD 0.1°C/°F; Resistance 0.01/0.1Ω
    Operating Relative Humidity 95% up to 30°C; 75% up to 40°C; 45% up to 50°C; 35% up to 55°C
    Maximum voltage 30V Between terminals or terminals and ground
    Operating altitude 3000 meters
    Dimensions 8" x 3.8" x 1.8" (205 x 98 x 46 mm) H x W x D
    Weight 17.7 ounces (475 grams)
    Power Requirements 6 type AAA batteries 1.5 V
    DC80R RTD Range and Accuracy Table
    Sensor Type Temperature Range Accuracy Input 4 Wire Accuracy Input 2/3 Wire Output Excitation Current mA
    Pt10 (385) -200 to 800°C
    -328 to 1472°F
    Not Specified Not Specified Not Specified 0.1 to 3.0
    Pt50 (385) -200 to 800°C
    -328 to 1472°F
    0.7 1.0 0.7 0.1 to 3.0
    Pt100 (385) -200 to 800°C
    -328 to 1472°F
    0.33 0.5 0.33 0.1 to 3.0
    Pt200 (385) -200 to 250°C
    -328 to 482°F
    0.2 0.3 0.2 0.1 to 3.0
    Pt200 (385) 250 to 630°C
    482 to 1166°F
    0.8 1.6 0.8 0.1 to 3.0
    Pt500 (385) -200 to 500°C
    -328 to 932°F
    0.3 0.5 0.3 0.05 to 0.8
    Pt500 (385) 500 to 630°C
    932 to 1166°F
    0.4 0.9 0.4 0.05 to 0.8
    Pt1000 (385) -200 to 100°C
    -328 to 212°F
    0.2 0.4 0.1 0.05 to 0.8
    Pt1000 (385) 100 to 630°C
    212 to 1166°F
    0.2 0.5 0.2 0.05 to 0.8
    Pt100 JIS -200 to 630°C
    -328 to 1166°F
    0.3 0.5 0.3 0.1 to 3.0
    The excitation current applies only to the simulate mode. The excitation current is provided by the target instrument (OHM meter of RTD meter) connected to the calibrator.
    DC80R Ohm Range and Accuracy Table
    Sensor Type Range Measurement accuracy (4 W ±Ω) Accuracy Simulate Ω Excitation Current mA
    Ohms 0 to 400Ω 0.1 0.15
    0.1
    0.1 to 0.5
    0.5 tp 3.0
    Ohms 400 to 1500.0Ω 0.5 0.5 0.05 to 0.8
    Ohms 1500.0 to 3200.0Ω 1
    2
    1 0.05 to 0.4
    The excitation current applies only to the simulation mode. The excitation current is provided by the target instrument (OHM meter of RTD meter) connected to the calibrator.
    Each DC80R includeds 6 size AAA batteries, one pair of test lead extension, one pair of stackable cable extension, one pair of heavy duty alligator clips, operation manual and carrying pouch.